DO30 series three-meter/multimeter calibrator
DO30-VI AC multifunctional calibrator
DO30-Ⅵ AC multifunctional calibrator is a multifunctional precision calibrator with multi-range, wide range, high-precision seven-and-a-half-digit output and display. The instrument can be displayed through a large color touch screen to facilitate manual detection by users. The instrument can output high-accuracy AC voltage/current, AC power, and AC power with power factor control. It is suitable for calibrating or verifying various AC voltmeters, AC ammeters, power meters, and power meters below level 0.1. Calibration devices such as factor tables, phase meters, frequency meters, and AC power analyzers can also be used as AC standard sources.
- Product description
-
I. Overview
DO30-Ⅵ AC multifunctional calibrator is a multifunctional precision calibrator with multi-range, wide range, high-precision seven-and-a-half-digit output and display. The instrument can be displayed through a large color touch screen to facilitate manual detection by users. The instrument can output high-accuracy AC voltage/current, AC power, and AC power with power factor control. It is suitable for calibrating or verifying various AC voltmeters, AC ammeters, power meters, and power meters below level 0.1. Calibration devices such as factor tables, phase meters, frequency meters, and AC power analyzers can also be used as AC standard sources. This instrument fully complies with the relevant requirements of JJG 124-2005 "Calibration Regulations for Ammeters, Voltmeters, Power Meters and Resistance Meters" and JJF 1587-2016 "Calibration Regulations for Digital Multimeters". It can be widely used in industrial and mining enterprises, measurement, aerospace, Quality inspection, measurement, teaching and research measurement standards and instrument maintenance work for military industry, scientific research institutes, railways, electric power, troops, schools and other units.
2. Main functions and featuresThe instruments that can be calibrated by this instrument are as follows: AC indicating instruments, AC voltmeters, AC ammeters, single-phase AC power meters, AC power analyzers, AC clamp meters, etc.
2.1 The maximum AC voltage output is 1100 V, the maximum current is 110A, and the AC frequency range is: 45Hz~1kHz.
2.2 It can accurately output AC active power, reactive power, apparent power, power factor, etc.
2.3 The maximum and minimum values of electricity can be automatically recorded through the software, the average value and standard deviation can be calculated, and the stability can be analyzed.
2.4 It has strong load capacity, ensuring stable and accurate values under full load conditions, and can cover the calibration of analog meters.
2.5 output has various value adjustment methods such as "fixed-point output, potentiometer adjustment, rotary encoder, step proportion" and so on.
2.6 Equipped with a large-size LCD touch color screen, the value display is intuitive and the operation is very convenient.
2.7 The output has complete software and hardware protection functions (overload protection, voltage short-circuit protection, current open-circuit protection, overheating protection, etc.) to improve the long-term reliability of the equipment.
2.8 Harmonics 2 to 31 times, amplitude 0 to 40%, phase fineness of each harmonic 0.010·N (N is the harmonic order)
3. Main technical indicatorsNote: The test environment temperature for the following indicators is 20℃±2℃.
3.1 AC voltage technical indicatorsMeasuring range resolution short term stability
(%RG/min)
Accuracy within one year
±(% of reading + % of full scale)
Distortion load capacity 45-200Hz
200-1000Hz
100mV
0.01mV
0.01
0.03%RD+0.02%RG
0.03%RD+0.02%RG
<0.2%
600mA
1V
0.01mV
0.005
0.02%RD+0.01%RG
30V
0.1mV
300V
1mV
150mA
1000V
10mV
0.01
0.03%RD+0.02%RG
50mA
3.2 AC current technical indicators
Measuring range resolution short term stability
(%RG/min)
Accuracy within one year
±(% of reading + % of full scale)
Distortion load capacity 45-200Hz
200-1000Hz
2mA
0.01μA
0.01
0.03%RD+0.02%RG
0.03%RD+0.02%RG
<0.2%
60V
10mA
0.1μA
100mA
1μA
0.01
60V
2A
0.01mA
20A
0.1mA
6V
50A
0.1mA
0.01
1.5V
100A
1mA
3.3 Technical indicators of power, power factor, phase and frequency
project scope Accuracy
AC power Combination of voltage range and current range(voltage ≥ 30V, current ≥ 0.01A)
Active power(|cosφ|≥0.5)
≤0.05%FS
Reactive power(|sin|φ≥0.5)
≤0.08%FS
Power factor/apparent power 1~-1 continuously adjustable <0.05%FS
Phase 0~360°
<0.02° (adjustment fineness 0.001°) frequency
(sine wave)
45Hz~200Hz
<0.02%
Adjustment fineness: 0.0001Hz 200Hz~600Hz
<0.03%
600Hz~1000Hz
<0.05%
AC power assessment scope:
Voltage: 75V, 150V, 300V, 600V; Current: 1A, 2A, 5A, 10A; Frequency: 50Hz. Other stalls available
Note: FS = voltage range and × current range.
3.4 Other technical indicatorsTemperature performance: Calibration temperature 18℃~22℃
Storage temperature -10℃~60℃
Temperature coefficient: The temperature exceeds the calibration temperature range. In the temperature range of 0℃~40℃, the temperature coefficient per degree Celsius is calculated as 20% of the technical index for one year.
Weight: ≤ 20kg (excluding accessories and packaging box)
Volume: 520mm×445mm×205mm
Working power supply: AC220V±10%
Working environment: temperature -20~40°C; relative humidity ≤85%.