DO30-Q type clamp meter calibrator


The DO30-Q clamp meter calibrator (hereinafter referred to as the calibrator) is an AC and DC clamp meter calibration device composed of an LED five-and-a-half-digit display AC and DC voltage and current calibration source and a clamp meter calibration coil. Amplitude adjustment can be controlled by keyboard (button) or potentiometer. It has the characteristics of digital display, wide measuring range, fine binning, high precision, good stability and easy use.

  • Product description
  • The DO30-Q clamp meter calibrator (hereinafter referred to as the calibrator) is an AC and DC clamp meter calibration device composed of an LED five-and-a-half-digit display AC and DC voltage and current calibration source and a clamp meter calibration coil. Amplitude adjustment can be controlled by keyboard (button) or potentiometer. It has the characteristics of digital display, wide measuring range, fine binning, high precision, good stability and easy use.
    Mainly used for calibration and testing of clamp ammeters, and suitable for calibrating and calibrating various current and voltmeters of level 0.2 and below. It can also be used as a high-stability test power supply, used in conjunction with high-level standard meters to calibrate ammeters and voltmeters above level 0.2. It is widely used in production, quality inspection, scientific research and instrument repair in industrial and mining enterprises, scientific research institutes, railways, electric power, troops, schools and other units.

    The main function
    1.1, 5 1/2 digit LED digital display output, which can be displayed in two ways: actual value or percentage.
    1.2. The AC and DC current output range is: 0~100uA~500uA~2mA~5mA~20mA~50mA~200mA~500mA~2A~5A~20A.
    1.3. The AC and DC voltage output range is: 0~200mV~1V~2V~5V~10V~20V~50V~100V~200V~500V~1000V.
    1.4. There are three AC frequencies available for selection: 50Hz, 60Hz, and 400Hz (use a crystal oscillator to ensure frequency stability).
    1.5. Output overload can be automatically protected and manually reset.
    1.6. There is a 10% margin for each voltage and current output range.
    1.7. Clamp meter measurement: Use our factory's calibration coil, which can measure 0~1000A; the conversion error is ±0.8%.

     
    Output items Output range Rated output maximum output Accuracy
    cross, straight 0~75mV~100mV~125mV~150mV~200mV~250mV 10mA 30mA  
    0~2.5V 20mA 60mA DC:
    0~5V~10V~25V 40mA 100mA ±(0.03% of reading
    0~250V~500V 20mA 60mA +0.02% of full scale)
    Current voltage 0~1000V 10mA 30mA AC(50Hz60Hz):
    cross, straight 0~50uA~60uA~75uA~90uA~100uA~500uA~2mA~5mA~20mA~50mA DC:3V   ±(0.05% of reading
    AC:36V +0.03% of full scale)
    flow current 0~200mA~500mA DC:3V    
    AC:12V  
    standard 0~2A~5A~22A 2V 3V  
    10、24、50、100、240、      
    500(Ω)1、2.4、5、10、24(kΩ)    
    resistance In addition, ×1000 and ×1 magnification gears (i.e. 10kΩ ~ 24MΩ) 0.25W ±0.2%+20mΩ
    Note: When the working environment temperature exceeds 23℃±2℃, for every 10℃ change, the additional error is less than the basic error of the range.
    Power consumption: AC power supply voltage 220V±10%, frequency 50Hz±1Hz; power consumption 150VA.
    Working environment: The temperature of the working environment is 5℃~35℃, and the relative humidity is﹤80%
    Working hours: Continuous
    Overall dimensions: width × height × depth 480 × 135 × 420 (mm)
    Weight: about 16kg