HG112 low resistance meter tester is a standard instrument for measuring, calibrating and verifying DC low resistance testers (digital bridge, microohmmeter, etc.). The HG112 low resistance meter tester has ten ranges: ×103Ω, ×102Ω, ×101Ω, ×100Ω, ×10-1Ω, ×10-2Ω, ×10-3Ω, ×10-4Ω, ×10-5Ω, ×10- 6Ω. Each range has 20 resistance values. The ×10-2Ω, ×10-3Ω, ×10-4Ω, ×10-5Ω, ×10-6Ω range is a dedicated DC analog resistance box, which uses a resistor network line and provides (0. 01mΩ~1μΩ) DC analog equivalent resistance.

  • Product description
  • HG112 low resistance meter tester is a standard instrument for measuring, calibrating and verifying DC low resistance testers (digital bridge, microohmmeter, etc.). The HG112 low resistance meter tester has ten ranges: ×103Ω, ×102Ω, ×101Ω, ×100Ω, ×10-1Ω, ×10-2Ω, ×10-3Ω, ×10-4Ω, ×10-5Ω, ×10- 6Ω. Each range has 20 resistance values. The ×10-2Ω, ×10-3Ω, ×10-4Ω, ×10-5Ω, ×10-6Ω range is a dedicated DC analog resistance box, which uses a resistor network line and provides (0. 01mΩ~1μΩ) DC analog equivalent resistance.

    This calibrator can calibrate DC resistance testers (digital bridge, microohmmeter) with a range of (1 μΩ~20kΩ) and an accuracy of ≤5×10-4.

    2. Technical indicators

         Technical indicators: ≤1×10-4~1×10-2
         Calibration temperature: 20℃±1℃, relative humidity: 40%~60%
         Operating temperature: 20℃±5℃, relative humidity: 25%~75%
         Size: 490×300×220
         Basic parameters: The basic parameters of each range of the calibrator are shown in the table below:

    Measuring range Adjustment range Step disc(Ω) level of accuracy Working power supply
    Reference value nominal value limit value

    1µΩ

    (0 20) µΩ

    1µΩ

    1

    15A

    20A

    25A

    10µΩ

    (0200) µΩ

    10µΩ

    0.1

    13A

    15A

    17A

    100µΩ

    (0  2) mΩ

    100µΩ

    0.1

    7A

    10A

    14A

    1mΩ

    (0 20) mΩ

    1mΩ

    0.02

    3A

    7A

    10A

    10mΩ

    (0200) mΩ

    10mΩ

    0.01

    2A

    4.5A

    5A

    100mΩ

    (0  2) Ω

    100mΩ

    0.01

    700mA

    1A

    1.4A

    (0 20) Ω

    0.01

    200mA

    200mA

    450mA

    10Ω

    (0200) Ω

    10Ω

    0.01

    70mA

    100mA

    140mA

    100Ω

    (0  2) kΩ

    100Ω

    0.01

    20mA

    20mA

    45mA

    1kΩ

    (0 20) kΩ

    1KΩ

    0.01

    7mA

    2mA

    14mA

     

    Used power: Low-range resistance disks ×10-2Ω, ×10-3Ω, ×10-4Ω, ×10-5Ω, ×10-6Ω are composed of resistor networks. The power of each plate should be different.

    Low-range resistance plate × 10-2Ω, rated current ≤ 2A

    Low range resistance plate ×10-3Ω, rated current ≤7A

    Low range resistance disk ×10-4Ω, rated current ≤10A

    Low range resistance disk ×10-5Ω, rated current ≤13A

    Low range resistance disk ×10-6Ω, rated current ≤15A

    Note: Please do not exceed the maximum current during use, and only for a short period of time, otherwise the resistor will be burned out.

    6. Insulation resistance: Under nominal conditions, the insulation resistance between the metal shell and the terminals of the instrument is not less than 500MΩ, and the test voltage is 500V.

    7. Withstand voltage test: Under nominal conditions, the instrument should withstand a 50Hz sine wave alternating current voltage of 1000V between the metal shell and the terminals, and the test lasts for 1 minute without breakdown or discharge.